Viswanadha Institute of Pharmaceutical Sciences
A rapid, sensitive, and accurate RP-HPLC method was developed for the identification and quantification of Acetaminophen and Oxycodone using a Waters PDA-detected HPLC system. Separation was achieved on an Inertsil ODS-C18 column (250 × 4.6 mm, 5 µm) with a flow rate of 1.0 mL/min. The mobile phase consisted of filtered and degassed Methanol and Acetonitrile in a 60:40 ratio. Detection was performed at a wavelength of 230 nm.
Percocet® is a combination analgesic containing oxycodone and acetaminophen. Under conditions of inflammation or hyperalgesia, opioid receptors in tissues can become upregulated and transported to nerve terminals. Acetaminophen (10-11) contributes to analgesia by inhibiting the COX-1 and COX-2 isoforms of cyclooxygenase, enzymes responsible for prostaglandin (PG) synthesis. Since prostaglandins are key mediators of pain sensation, reducing their production increases the pain threshold.
Oxycodone (8-9) and its active metabolites selectively bind to the μ-opioid receptor, and to a lesser extent the κ and δ receptors, in both the central nervous system and peripheral tissues. This interaction activates G-protein–coupled receptor signaling. Activation of μ-opioid receptors inhibits N-type voltage-operated calcium channels, reducing neurotransmitter release and thereby diminishing the perception of pain.
II. MATERIALS AND METHODS
Preparation of Stock solution:
100 mg of Oxycodone and 100 mg of Acetaminophen API standards were accurately weighed and are transferred into two separate 100 ml volumetric flasks, dissolved in mobile phase, then sonicated for 20 minutes to obtain 1000μg/ml.
Preparation of working standard solution:
From the above standard stock solution, 4 ml from each solution was transferred into 100ml volumetric flasks, made up to the volume with mobile phase to get 40μg/ml of Oxycodone and 40μg/ml of Acetaminophen.
III. RESULTS AND DISCUSSION
Method validation:
Validation parameters include specificity, linearity, range, accuracy, precision, limit of detection, limit of quantification, robustness and assay (1-7).
Specificity:
Specificity is the ability to assessing equivocally the analyte in the presence of components which may be expected to be present. Typically these components include impurities, degradants, matrix etc. Blank solution and standard solutions of Oxycodone(40μg/ml) and Acetaminophen(40μg/ ml) were injected into the HPLC system. The peak purity data of Oxycodone and Acetaminophen were compared. There should not be any interference at the retention time of the main peaks.
Linearity:
Linearity for the drugs Oxycodone and Acetaminophen (12-19) was determined by preparing the standard solutions at six concentrations levels in the range of 20-70μg/ml Oxycodone and 20-70μg/ml for Acetaminophen from stock solution. The linearity charts of Oxycodone and Acetaminophen was shown in the figure no 2&3. The correlation coefficient was found to be 0.9997 and 0.9993 for Oxycodone and Acetaminophen respectively. Linearity results were tabulated in table 2.
Accuracy:
Accuracy was performed by spiking known amounts of standard solution to sample solution at three different concentrations levels (50%, 100%, 150%) and there by analyzed for %RSD which should not be more than 2.0.The % recovery was calculated and the results was reported in table no. 3 & 4.
Precision:
The precision (7-12) of the analytical method was studied by injecting six replicates of standard containing 40μg/ml of Oxycodone and 40μg/ml of Acetaminophen which were injected into HPLC system. The % RSD was calculated and the results were reported in the table no.5 & 6.
Limit of Detection (LOD) and Limit of Quantification (LOQ):
The limit of detection was defined as the concentration which yields a signal - to – noise ratio 3:1 whereas the limit of quantification was calculated to be the lowest concentration that could be measured with signal - to – noise ratio10:1. LOD and LOQ were calculated from slope and standard deviation. The results were tabulated in table no. 7.
Robustness:
The smallest deliberate changes in method like change in flow rate are made but there were no predictable changes in the results and are in the range as per ICH guidelines. Conditions like decrease in flow rate (0.8 ml/min), increase in flow rate (1.2 ml/min) was maintained and samples were injected in duplicate manner. System suitability parameters were not much affected and all the parameters were passed. % RSD was found to be within the limits and results were tabulated in table no. 8.
Assay:
Assay was conducted on marketed formulation and mean % assay was found. The results were tabulated in table no. 9.
Table1: Optimised Chromatographic conditions
|
Parameters |
Method |
|
Stationary Phase(column) |
Inertsil -ODS C18(250 x 4.6 mm, 5 µ) |
|
Mobile Phase |
Methanol and Acetonitrile (60:40) |
|
Flow rate (ml/min) |
1.0 ml/min |
|
Run time(minutes) |
8 min |
|
Temperature in the column (°C) |
Ambient |
|
Volume of injection loop (ml) |
20 |
|
Wavelength of detection (nm) |
230nm |
|
Drug RT (min) |
3.570 min for Oxycodone and 5.297 for Acetaminophen. |
Figure 1: Optimized chromatogram
Table 2: Linearity data of Oxycodone and Acetaminophen
|
Oxycodone |
Acetaminophen |
||
|
Conc (µg/ml) |
Peak area |
Conc (µg/ml) |
Peak area |
|
20 |
534254 |
20 |
983353 |
|
30 |
775469 |
30 |
1498937 |
|
40 |
1025584 |
40 |
2017341 |
|
50 |
1275586 |
50 |
2629425 |
|
60 |
1567986 |
60 |
3138733 |
|
70 |
1817698 |
70 |
3645876 |
Figure 2: Calibration Curve of Oxycodone
Figure 3: Calibration Curve of Acetaminophen
Table 3: Accuracy data of Oxycodone
|
Concentration % of spiked level |
Amount added (ppm) |
Amount found (ppm) |
% Recovery |
Statistical Analysis of % Recovery |
|
50% - 1 |
20 |
19.92 |
99.78 |
99.87 |
|
50% - 2 |
20 |
19.96 |
99.89 |
|
|
50% - 3 |
20 |
19.88 |
99.78 |
0.68 |
|
100 % - 1 |
40 |
39.94 |
98.92 |
99.84 |
|
100 % - 2 |
40 |
39.86 |
99.75 |
|
|
100% - 3 |
40 |
39.76 |
99.08 |
0.657 |
|
150% - 1 |
60 |
59.97 |
99.96 |
100.07 |
|
150% - 2 |
60 |
60.02 |
100.08 |
|
|
150% - 3 |
60 |
60.01 |
100.01 |
0.345 |
Table 4: Accuracy data of Acetaminophen
|
Concentration % of spiked level |
Amount added (ppm) |
Amount found (ppm) |
% Recovery |
Statistical Analysis of % Recovery |
|
50% - 1 |
20 |
19.89 |
99.87 |
99.97 |
|
50% - 2 |
20 |
19.72 |
99.87 |
|
|
50% - 3 |
20 |
20.08 |
100.03 |
0.874 |
|
100 % - 1 |
40 |
39.92 |
99.88 |
99.97 |
|
100 % - 2 |
40 |
40.01 |
100.07 |
|
|
100% - 3 |
40 |
40.05 |
100.08 |
0.687 |
|
150% - 1 |
60 |
59.95 |
98.87 |
99.94 |
|
150% - 2 |
60 |
59.97 |
99.88 |
|
|
150% - 3 |
60 |
59.98 |
99.98 |
0.97 |
Table 5: System Precision data of Oxycodone and Acetaminophen
|
S. No |
Peak areas of Oxycodone |
Peak areas of Acetaminophen |
|
1 |
1025863 |
2035142 |
|
2 |
1025642 |
2038624 |
|
3 |
1025784 |
2039586 |
|
4 |
1025537 |
2038867 |
|
5 |
1025682 |
2037695 |
|
Mean |
1025702 |
2037983 |
|
SD |
126.3618 |
1725.99 |
|
% RSD |
0.01232 |
0.084691 |
Table 6: Method Precision data of Oxycodone and Acetaminophen
|
S. No |
Peak areas of Oxycodone |
Peak areas of Acetaminophen |
|
1 |
1025867 |
2037861 |
|
2 |
1028965 |
2038967 |
|
3 |
1029958 |
2037765 |
|
4 |
1029685 |
2039685 |
|
5 |
1028968 |
2038567 |
|
6 |
1028567 |
2038868 |
|
Mean |
1028668 |
2038619 |
|
SD |
1465.087 |
724.713 |
|
% RSD |
0.142426 |
0.035549 |
Table 7: LOD and LOQ data of Oxycodone and Acetaminophen
|
Drug Name |
LOD (µg/ml) |
LOQ (µg/ml) |
|
Oxycodone |
0.65 |
1.97 |
|
Acetaminophen |
0.27 |
0.82 |
Table 8: Robustness data of Oxycodone and Acetaminophen
|
Sr No |
Drug Name |
Condition |
Peak area |
% RSD |
|
1 |
Oxycodone |
Decreased Flow rate of 0.8 ml/min |
1016744 |
0.106 |
|
2 |
Increased Flow rate of 1.2 ml/min |
1038506 |
0.132 |
|
|
3 |
Acetaminophen |
Decreased Flow rate of 0.8 ml/min |
2026242 |
0.156 |
|
4 |
Increased Flow rate of 1.2 ml/min |
2046930 |
0.085 |
Table 9: Assay data Oxycodone and Acetaminophen
|
Sr. No |
Peak area of Oxycodone |
% Assay |
Peak area of Acetaminophen |
% Assay |
|
1 |
1025863 |
98.96 |
2035142 |
98.82 |
|
2 |
1025642 |
2038624 |
||
|
3 |
1025784 |
2039586 |
||
|
4 |
1025537 |
2038867 |
||
|
5 |
1025682 |
2037695 |
CONCLUSION
The developed RP-HPLC method was validated as per ICH guidelines. All the system suitability parameters were within the range as stated by ICH guidelines. Interference peaks were not observed in blank, standard and sample chromatogram. Hence simple, precise and accurate, sensitive, specific and robust method was developed and validated. This can be used in quality control department with respect to routine analysis.
ACKNOWLEDGEMENTS:
Authors are thankful to the management of Viswanadha Institute of Pharmaceutical Sciences (VNIPS) for providing facilities and support to carry out this work.
REFERENCE
P Surekha, Dr. P V Madhavi Latha, K Suvarna, Dr. P Uma Devi, Method Development and Validation of Simultaneous Estimation of Oxycodone and Acetaminophen in Bulk and Pharmaceutical Dosage Forms by RP-HPLC Method, Int. J. of Pharm. Sci., 2025, Vol 3, Issue 11, 4438-4444. https://doi.org/10.5281/zenodo.17735856
10.5281/zenodo.17735856