Noble Pharmacy College, Noble University,"Parth-Vatika", Junagadh, Gujarat, India. 362310
A simple, precise, accurate, and robust RP-HPLC method was developed and validated for the quantitative estimation of galantamine in tablet dosage form. The chromatographic separation was achieved on a [insert column type, e.g., C18 column (250 mm × 4.6 mm, 5 µm particle size)] using a mobile phase consisting of [e.g., methanol:water or acetonitrile:buffer] in the ratio [e.g., 70:30 v/v], at a flow rate of [e.g., 1.0 mL/min]. The detection was carried out at [e.g., 288 nm] using a UV detector. The method was validated as per ICH Q2(R1) guidelines for system suitability, linearity, accuracy, precision, robustness, and specificity. The method showed good linearity in the concentration range of [e.g., 10–100 µg/mL] with correlation coefficient (R²) > 0.999. The developed method was successfully applied for the estimation of galantamine in marketed tablet formulations
What is Dementia
Dementia is a general term for a group of conditions characterized by a decline in cognitive functioning severe enough to interfere with daily life and independence1. It primarily affects memory, thinking, reasoning, and socialabilities. Alzheimer'sdisease isthe most commoncause of dementia, followed by vascular dementia and other types2.
What are the Different types of dementia?
Types of dementia include:
Mixed dementia, a combination of two or more types of dementia. For example, through autopsy studies involving older adults who had dementia, researchers have identified that many people had a combination of brain changes associated with different forms of dementia.
Basic Principles of HPLC
HPLC works on the principle of differential migration of analytes based on their affinity for the stationary and mobile phases. The process can be summarized as follows:
Components of method validation: The following are typical analytical performance characteristics which may be tested during methods validation:[15-17]
Accuracy
Accuracy is defined as the nearness of a measured value to the true or accepted value. Practically accuracy indicates the deviation between the mean value foundand the true value. It is determined by applying the method to samples to which known amounts of analyte have been added. These should be analysed against standard and blank solutions to ensure that no interference exists. The accuracy is then calculated from the test results as a percentage of the analyte recovered by the assay. It may often be expressed as the recovery by the assay of known, addedamounts of analyte.
Precision
It expresses closeness of agreement (degree of scatter) between a series of measurements obtained from multiple sampling of the same homogeneous sample under the prescribed conditions. Precision is a measure of the reproducibility of the whole analytical method. It consists of two components: repeatabilityand intermediate precision. Repeatabilityisthe variationexperienced by a single analyst on a single instrument. It does not distinguish between variation from the instrument or system alone and from the sample preparation process. During validation, repeatability is performed by analysing multiple replicates of an assaycomposite sample by using the analytical method. The recovery value is calculated. Intermediate precision is the variation within a laboratory such as different days, with different instruments, and by different analysts. Accuracyandprecision are not the same, as the diagram below indicates. A method can have good precision and yet not be accurate.
Linearity
Linearity is the ability of analytical procedure to obtain a response that is directlyproportional to the concentration (amount) of analyte in the sample. If the methodis linear, the test results are directly or by well-defined mathematicaltransformation proportional to concentration ofanalyte in samples within a givenrange. Linearity is usually expressed as the confidence limit around the slope of the regression line.
Limits of detection and quantitation:
The limit of detection (LOD) is defined as the lowest concentration of an analyte in a sample that can be detected, not quantified. LOD is expressed as a concentration at a specified signal: noise ratio, usually 3:1. The limit of quantitation (LOQ) is defined as the lowest concentration of an analyteinasamplethat can be determined with acceptable precision and accuracyunder the stated operational conditions of the method. For LOQ, ICH has recommended a signal: noise ratio 10:1. LOD and LOQ may also be calculated based on the standard deviation of the response (SD) and the slope ofthe calibrationcurve(s) at levels approximatingthe LOD according tothe givenbelow formulae.
????????????=3.3×???? /????????
????????????=10×???? /????????
Specificity
Specificity is the ability to assess unequivocally the analyte in the presence of components which maybe expected to be present. Typically, these might includeimpurities, degradants, matrix, etc. Lack of specificityof an individual analytical procedure may be compensated byother supporting analyticalprocedure(s). Thisdefinition has the following implications: Identification: to ensure the identityof an analyte. PurityTests: to ensure that all the analyticalprocedures performed allow an accurate statement of the content of impurities of an analyte, i.e. related substances test, heavy metals, residualsolventscontent,etc. Assay(content or potency): to provide anexact result which allows an accurate statement on the content or potency of the analyte in a sample.
Range
The range ofthe method is the interval between the upper and lower levels ofan analyte that have been determined with acceptable precision, accuracy and linearity. It is determined on either a linear ornonlinear response curve (iewheremorethanone range is involved, asshownbelow) and is normally expressed in the same units as the test results.
Robustness:
The robustness of an analytical procedure is a measure of its capacity to remain unaffected by small, but deliberate variations in method parameters and provides an indication of its reliability during normal usage.
Validation parameter
Parameters |
Acceptance criteria |
Linearity |
Correlation coefficient r2>0.999 |
Accuracy |
Recovery 98-102% (individual) |
Precision |
RSD<2% |
Repeatability |
RSD<2% |
Intermediate Precision |
RSD<2% |
Specificity/ Selectivity |
Nointerference |
Range |
80–120% |
Detection Limit (DL) |
S/N>2or3 |
AIM
The aim of the present work is to develop and validate a precise, accurate, and robust RP-HPLC method for the quantification of Galantamine in tablet dosage form.
OBJECTIVE
MATERIALS AND METHODS
Table: Chemical and Reagents Used
Reagent |
Purpose |
Source |
Galantamine (API) |
Active Pharmaceutical Ingredient for calibration and sample analysis |
Commercial Supplier |
Galantamine Tablets |
Commercial formulation for Sample analysis |
Sun Pharma |
Methanol |
Solvent for preparation of standard and sample solutions |
Merck (HPLC Grade) |
Water |
Solvent for mobile phase and dilutions |
HPLC Grade, Merck |
Acetonitrile |
Solvent for mobile phase and sample preparation |
Merck (HPLC Grade) |
Buffer Solution |
To control pH of mobile phase(optional) |
Merck, pH6.4 |
Table: Instruments Used
Instrument/Equipment |
Purpose |
Specification |
Source |
HPLC System |
Separation and quantification of Galantamine |
UV Detector, Pump, Injector |
Agilent, Shimadzu, Waters |
C18 Column (250mmx 46 mm, 5 µm) |
Stationary phase for chromatographic separation |
C18 Silica |
Waters, Agilent |
UV Detector |
Detection of Galantamine at 210nm |
UV-Visible Detector |
Agilent, Shimadzu |
Analytical Balance |
Weighing Galantamine and tablet powder |
0.1mg accuracy |
Mettler Toledo, Shimadzu |
Sonicator |
For dissolving and preparing samples |
40kHz frequency |
Branson, Labsonic |
pH Meter |
To measure pH for mobile phase preparation |
pH range: 0–14 |
Thermo Fisher |
Glassware (Volumetric Flasks, Pipettes, etc.) |
To prepare standard and sample solutions |
Standard laboratory equipment |
Borosil, Kimble |
Table: Preparation of Standard and Sample Solutions
Solution |
Preparation Method |
Concentration |
Volume |
Purpose |
Standard Stock Solution |
Weigh 10 mg of Galantamine, dissolve in methanol, and dilute to volume with methanol. |
100 µg/mL |
100mL |
Calibration standard |
Working Standard Solution |
Dilute standard stock solution to desired concentration with mobile phase. |
1–50 µg/mL |
As required |
For calibration curve preparation |
Sample Solution |
Weigh 10 tablets, grind to powder, dissolve in 10 mL methanol, sonicate, and dilute to 100 mL with mobile phase. |
10 µg/mL |
10mL |
Sample analysis |
Table: Chromatographic Conditions
Parameter |
Condition |
Stationary Phase |
C18 Column (250mm×4.6mm,5µm) |
Mobile Phase |
Methanol:Water (60:40v/v) |
Flow Rate |
1.0mL/min |
Injection Volume |
20µL |
Detection Wavelength |
210nm |
Column Temperature |
Ambient(20–25°C) |
Run Time |
10 minutes |
Table: Method Validation Parameters
Validation Parameter |
Procedure |
Acceptance Criteria |
Specificity |
Inject blank, standard, and sample solutions; check for interference. |
No interference at Galantamine’s retention time (~6.5 min). |
Linearity |
Prepare a series of standard solutions, plot the calibration curve. |
R2>0.99R^2>0.99R2>0.99 |
Accuracy (Recovery) |
Add known amounts of Galantamine to the sample and calculate the recovery. |
98–102% recovery at 80%, 100%,120% spiked levels. |
Precision (Intraday) |
Inject the same sample repeatedly within the same day (6 times). |
%RSD≤2% |
Precision (Interday) |
Inject the same sample on different days (3 days). |
%RSD≤2% |
Repeatability |
Inject the same sample multiple times (6 times) to check reproducibility. |
%RSD≤2% |
LOD & LOQ |
Calculate the signal-to-noise ratio and determine the lowest detectable and quantifiable concentration. |
LOD ≤ 0.05 µg/mL, LOQ ≤ 0.15 µg/mL |
Robustness |
Deliberately change parameters like flow rate and mobile phase composition. |
No significant effect on retention time or peak symmetry. |
Fig: Structure of Galantamine
Chromatographic condition
The chromatographic separation of Galantamine was achieved on C-18 (id4.6x250mm, 5 µm) by using mobile phase composed of Methanol: Water (60:40 v/v %), at flow rate 1.0 ml/min with run time of 10 minutes. Detection of drug was carried out at 210 nm by using diluent as mobile phase.
Method of validation
As per ICH guideline (Q2R1), the method validation parameters studied were specificity, linearity, accuracy, precision, limit of detection, limit of quantitation and robustness.
A Specificity
The analytical method for specificity was evaluated by injecting the following solutions. Diluent was prepared and inject into the HPLC system in triplicate. Sample solution was prepared with appropriate levels of excipients as a placebo sample and inject into the HPLC system in triplicate for all the dosage strengths. Placebo was prepared by mixing all excipients without active ingredients. Standard and sample solutions were prepared for assay(100% Conc.) and inject into the HPLC system in triplicate.
A Linearity and Range
Preparation of Solution for linearity studies: For the purpose of linearity, accurately weighed amount of Galantamine (10 mg) was taken into the volumetric flask (10 ml) and volume of the flask was raised to 10 ml with methyl alcohol to give stock solution containing 100 µg/ml of Galantamine. Various aliquots from this stock solution were transferred to another 10 ml volumetric flask and volume was raised to the mark with mobile phase to give final solutions containing 2, 4, 6, 8 and 10 µg/ml of Galantamine respectively.
Precision
Repeatabiliy
Prepared standard working solution of mixtures having concentration of Galantamine (4 μg/ml) was injected at volume of 20 μL into column by employing optimized chromatographic conditions. Each standard mixture was injected 5 time and peak area was monitored. Each concentration was monitored for repeatability by RSD. Intra-day and Inter-day Precision
System Suitability Parameters
Solution of Galantamine (15 μg.ml-1) was injected 3 times for determination of System suitability parameters which includes Retention time (Rt), Tailing factor (Tf), Resolution (Rs) and number of theoretical plates. System suitability parameters for selected concentration were determined by C.V.
Accuracy
Accuracy of the analytical method has been performed by spiking of sample with the standard. Spiking of the placebo was performed at 50,100 and 150 % of the target concentration
Limit of detection and Limit of Quantification
The limit of detection (LOD) and the limit of quantification (LOQ) were calculated using the standard deviation of y-intercept of calibration curve. The limit of detection (LOD) and the limit of quantification (LOQ):
LOQ=10σ/s and LOD=3.3σ/s
Where, σ = the standard deviation of the response.
S =the slope of the calibration curve
Robustness
Following parameters were alteredone byone for determinationofrobustness ofthe method and their effect was observed by comparing with the standard preparation. Mobile phase flowrate (± 0.1mL/min),optimizedflowratewas1.0mL/min.Mobilephasecomposition(±2mL),in optimized ratio2 determinations of Galantamine= 2 µg/mL for each alteration were carried outand RSD was measured
Result and Discussion
Selection of Wavelength
To determine wavelength for measurement, standard spectra of Galantaminewas scannedbetween 200-400 nm against diluents. Absorbance maxima of Galantamine have detected at 210nm. Chromatogramwas taken at 219 nm, druggive good peak height and shape. So, 210 nm was selected for estimation of Galantaminein formulation.
Selection of Mobile phase
Trail 1
Fig: Trial 1: Chromatogram of Galantamine (15µg,ml-1)
Trail 2
Fig:Trial 2:Chromatogram of Galantamine (15 µg,ml-1)
Trial 3
Fig: Trial 3: Chromatogram of Galantamine (15 µg,ml-1)
Table 6.10 Mobile phase Trial for Galantamine
Trial |
Mobile Phase |
Ratio |
Result |
1 |
Acetonitrile: Water |
(30:70v/v) |
No peak detected |
2 |
Methanol: Water |
(50:50v/v) |
Peaks detected and separated, but broad peaks observe. |
3 |
Methanol: Water |
(60:40v/v) |
Good peak with Adequate solution was observed. |
Chromatographic conditions for optimized mobile phase trial
Fig6.6: Optimized mobile phase trial for optimized chromatogram of Std. Galantamine: 4.315min
Fig : Chromatogram of blank Methanol: Water (60:40v/v)
Method Validation
Linearity
For the purpose of linearity, accurately weighed amount of Galantamine(10 mg)was taken into the volumetric flask (10 ml) and volume of the flask was raised to 10 ml with methyl alcohol to give stock solution containing 100 µg/ml ofGalantamine. Various aliquots from this stock solution were transferred to another 10 ml volumetric flask and volume was raised to the mark with mobile phase to give final solutions containing 2, 4, 6, 8 and 10µg/ml of Galantamine
Table: Linearity data for Galantamine
|
Galantamine |
||
Conc. (µg/ml) |
Mean Area |
±SD(n=5) |
% RSD |
5.0 |
382534 |
382534.3±1761.43 |
0.46 |
10.0 |
612609 |
612609±1409.31 |
0.23 |
15.0 |
764730 |
764730.7±1240.80 |
0.16 |
20.0 |
921136 |
921136.3± 2446.21 |
0.16 |
25.0 |
1117955 |
1117955±5261.42 |
0.47 |
Fig: Overlain Linearity Spectra of Galantamine
Fig : Calibration curve of Galantamine
Table: Linearity results for Galantamine
Regression Analysis |
Galantamine |
Concentration Range |
5-25μg/mL |
Regression equation |
y = 35708x + 225420 |
Correlation co-efficient |
0.995 |
Precision
Repeatability
The data for repeatability for Galantamine is shown in table 6.13. The %R.S.D For Repeatability data was found to be 1.10 % for Galantamine.
Table: Repeatability data for Galantamine
Drugs |
Conc. (µg/ml) |
Mean Peak Area ± SD |
%RSD |
Galantamine |
15 |
724860 ± 1041.54 |
1.10 |
Inter-day precision
The data for interday precision for Ranitidine and Ondansetron is shown in table 6.14. The % R.S.D for intraday precision was found to be 0.33-0.63% for Galantamine.
Table: Inter-day precision data for estimation of Ranitidine and Ondansetron
|
Galantamine |
||
(µg/ml) |
5 |
15 |
25 |
|
387658 |
763456 |
926578 |
|
387689 |
768790 |
923673 |
|
383434 |
766589 |
929803 |
MEAN |
386260.3333 |
766276.3333 |
926684.6667 |
±SD |
2446.725543 |
2680.536203 |
3066.391745 |
RSD |
0.633698398 |
0.349812344 |
0.330899156 |
Intra -day precision
The data for intra-day precision for Galantamine is shown in table 6.15. The % R.S.D for intraday precision was found to be 0.35-0.57% for Galantamine.
Table: Intra-day precision data for estimation of Galantamine
|
Galantamine |
||
(µg/ml) |
5 |
15 |
25 |
|
385467 |
768790 |
926589 |
|
385089 |
763456 |
922671 |
|
381452 |
764980 |
927689 |
MEAN |
384002.6667 |
765742 |
925649.6667 |
±SD |
2216.01293 |
2746.430072 |
2636.582479 |
RSD |
0.577343108 |
0.35879318 |
0.284943924 |
Accuracy
Accuracy of the method was confirmed by recovery study from synthetic mixture at three level standard additions. Percentage recovery for Galantaminewas found to be 99.48- 99.78%. The results are shown in table.6.16.
Table: Recovery data for Galantamine
|
Galantamine |
|||||
|
50% |
100% |
150% |
|||
|
Amount of drug recovered (mg) |
% Recovery |
Amount of drug recovered (mg) |
% Recovery |
Amount of drug recovered (mg) |
% Recovery |
|
1.46 |
99.76 |
2.97 |
99.20 |
4.54 |
100.20 |
|
1.40 |
96.70 |
2.89 |
99.01 |
4.56 |
100.22 |
|
1.56 |
100.50 |
3.09 |
100.01 |
4.68 |
100.30 |
Mean |
1.49 |
96.65 |
2.98 |
99.43 |
4.69 |
100.24 |
%RSD |
0.02 |
1.30 |
0.04 |
1.75 |
0.05 |
0.68 |
LOD and LOQ
The limit of detection (LOD) and Limit of Quantification (LOQ) was found to be as per below:
Table LOD and LOQ Limit for Galantamine
Galantamine |
|
LOD(μg/ml) |
LOQ(μg/ml) |
3 |
10 |
Selectivity
There is no interference in the mixture.
Robustness
The method is found to be robust as the results were not significantly affected by slight variation in Mobile Phase Composition and flow rate of mobile phase. The results are shown in table 6.19. Variation seen was within the acceptable range respect to peak asymmetry and theoretical plates, so the method was found to be robust.
Table: Robustness data for Galantamine
Parameter |
Level of Change |
Effect on assay volume |
|
Galantamine |
|||
Assay ± SD |
%RSD |
||
Flow rate |
0.9 mL/min |
95.70 ±0.50 |
0.49 |
1.1 mL/min |
101.09 ±0.72 |
0.72 |
|
Mobile phase composition |
50:50 |
95.47 ±0.53 |
0.53 |
60:40 |
95.39 ±0.99 |
0.98 |
|
30:70 |
99.51 ±0.67 |
0.67 |
Analysis of marketed product
The proposed method was successfully applied to analysis of the commercially available tablet formulation. The % drugs were found satisfactory, which is comparable with the corresponding label claim.
Table: Analysis of marketed formulations
Drug |
Amount taken (µg/mL) |
Amount found (µg/mL) |
% Assy |
Ranitidine |
3 |
2.93±0.04 |
99.80±1.20 |
SUMMARY OF METHOD VALIDATION
Table: Summary of validation parameter of RP-HPLC method
Optimized chromatographic Condition |
|
Stationary Phase |
C-18 (id4. 6x250 mm,5 µm) |
Mobile Phase |
Methanol: Water(60:40v/v) |
Detection wavelength |
210nm |
Flow rate |
1 ml/minute |
Run time |
10 minutes |
Retention Time |
4.315min |
Validation parameters |
|||
Parameter |
Limit |
Result |
Conclusion |
Galantamine |
|||
Linearity and Range |
R2>0.995 |
0.9992 (2-10µg/mL) |
Method was linear |
Repeatability |
RSD<2 |
1.10 |
Method was repeatable |
LOD |
- |
3 |
- |
LOQ |
- |
10 |
- |
Intra-day Precision |
RSD<2 |
0.33-0.63 |
Method was precise |
Inter-Day Precision |
RSD<2 |
0.35-0.57 |
Method was precise |
%Recovery |
98-102% |
99.35± 0.83 – 100.01± 0.03% |
Method was accurate |
Robustness |
RSD<2 |
0.41 – 0.63 |
Method was robust |
Assay% |
|
99.80 ± 1.20 |
- |
REFERENCES
Marutkumar Parmar, Dhirendra Kumar Tarai, Khyati Bhupta, Dr. Santosh Kirtane, Development And Validation of RP HPLC Method for Estimation of Galantamine in Tablet Dosage Form, Int. J. of Pharm. Sci., 2025, Vol 3, Issue 6, 3179-3191. https://doi.org/10.5281/zenodo.15716476