View Article

Abstract

Instrument calibration is one of the primary processes used to maintain instrument accuracy. Calibration is the process of configuring an instrument to provide a result for a sample within an acceptable range. There are three main reasons for having instruments calibrated are to ensure readings from an instrument are consistent with other measurements, to determine the accuracy of the instrument readings, to establish the reliability of the instrument i.e. that it can be trusted. This review includes the information about the tests conducted for calibrating different analytical instruments and acceptance criteria. Out of calibration is the major thing during analysis. It gives in detail about the out of calibration also.

Keywords

pH meter, weighing balance-performance check-out of calibration

Introduction

Calibration is a comparison between measurements one of known magnitude or correctness made or set with one device and another measurement made in as similar a way as possible with a second device. The device with the known or assigned correctness is called the standard. The second device is the unit under test, test instrument, or any of several other names for the device being calibrated. The formal definition of calibration by the International Bureau of Weights and Measures is the following: "Operation that, under specified conditions, in a first step, establishes a relation between the quantity values with measurement uncertainties provided by measurement standards and corresponding indications with associated measurement uncertainties (of the calibrated instrument or secondary standard) and, in a second step, uses this information to establish a relation for obtaining a measurement result from an indication.

Calibration may be called for

  1. a new instrument
  2. after an instrument has been repaired or modifie
  3. when a specified time period has elapsed
  4. when a specified usage (operating hours) has elapsed
  5. before and/or after a critical measurement
  6. after an event,  for example  after an instrument has had a shock, vibration, or has been exposed to   an   adverse condition which potentially may have put it out of calibration or damage it
  7. sudden changes in weather
  8. whenever observations appear questionable or instrument indications do not match the output of surrogate instruments
  9. as specified by a requirement, e.g., customer specification, instrument manufacturer recommendation.

In general use, calibration is often regarded as including the process of adjusting the output or indication on a measurement instrument to agree with value of the applied standard, within a specified accuracy.  For example, a thermometer could be calibrated so the error of indication or the correction is determined, and adjusted (e.g. via calibration constants) so that it shows the true temperature in Celsius at specific points on the scale. This is the perception of the instrument's end-user. However, very few instruments can be adjusted to exactly match the standards they are compared to. For the vast majority of calibrations, the calibration process is actually the comparison of an unknown to a known and recording the results.

CALIBRATON:

  1. Calibration of an instrument is the process of determining its accuracy. The process involves obtaining a reading from the instrument and measuring its variation from the reading obtained from a standard instrument.
  2. Calibration of an instrument also involves adjusting its precision and accuracy, So that its reading comes in accordance with the established standard.
  3. This is important for justifying the process of Qualification and Validation.
  4. The instrument or equipment with the known accuracy is known as standards. All the other instruments are measured against this standard.
  5. It is important to know that the standards vary from one country to the other depending upon the type of industry.

OBJECTIVE:

  1. Ensure measurement accuracy – Confirm that the instrument provides results within acceptable limits.
  2. Maintain consistency – Ensure that measurements remain consistent over time and across different instruments.
  3. Meet regulatory or industry standards – Comply with legal, quality, or safety requirements.
  4. Detect and correct errors – Identify deviations or drifts in performance and correct them.
  5. Ensure product quality – Support quality control in manufacturing, testing, and research by using precise measurements.

CALIBRATION TECHNIQUES:

1. Direct Calibration : Compares the instrument directly with a reference standard.Simple and accurate if the reference is traceable.

Example: Using a certified thermometer to check another thermometer’s reading.

2. Indirect Calibration : Involves calibrating through an intermediate measurement or another system. Often used when direct comparison is not possible.

Example: Calibrating a flow meter by measuring the output volume over time

3. Zero and Span Calibration  :

Zero: Adjusts the device to show a correct reading when the input is zero.

Span: Adjusts to ensure accurate readings across the range. Common in pressure, flow, and temperature sensors.

4. Multipoint Calibration : Calibration at several points across the instrument's full range.

Improves accuracy and detects non-linear behavior. Often used for sensors and analytical instruments.

5. One-point Calibration : Calibration at a single known value. Simple and quick but less accurate for wide-range instruments.

6. Two-point Calibration : Uses two reference values (often low and high).Common for linear systems, balancing simplicity and accuracy.

7. Dynamic Calibration : Applies to systems where conditions change over time (e.g., acceleration, vibration).Measures performance under real-time, changing conditions.

 8. Automatic or Self-Calibration : Instruments with built-in functions to calibrate themselves. Common in digital devices and automated systems.

9. Field Calibration (In-situ) : Calibration performed at the location where the device is used. Reduces downtime but may be less precise than lab calibration.

10. Laboratory Calibration : Conducted in a controlled environment using high-precision standards. Provides the highest accuracy and traceability.

CLEANING & CALIBRATION OF PH METER:

1. Introduction :

The accurate measurement of pH is crucial for ensuring the quality and compliance of various products and processes. A pH meter requires regular cleaning and calibration to maintain accuracy and reliability.

2. Objectives :

  • To ensure the pH meter is free from contaminants that could affect its performance.
  • To calibrate the pH meter using standard buffer solutions for accurate measurements.

3. Equipment and Materials :

  • pH meter
  • Standard buffer solutions (pH 4.00, 7.00)
  • Distilled water
  • Beakers
  • Tissue

4. Cleaning Procedure:

Preparation:

  • Gather all necessary materials and ensure a clean working area.
  • Wash hands and wear appropriate personal protective equipment (PPE).

Cleaning the pH Meter :

  • Rinse the pH electrode with distilled water to remove any residual samples.
  • If the electrode is contaminated, soak it in a suitable cleaning solution as per the manufacturer's recommendations.
  • Gently wipe the electrode with a soft lint-free cloth to remove any stubborn residues.
  • Rinse the electrode again with distilled water.

Inspecting the Electrode :

  • Check the electrode for any signs of damage or wear.
  • Ensure that the reference junction is clean and functioning properly.

5. Calibration Procedure:

Preparation:

  • Turn on the pH meter and allow it to warm up as recommended by the manufacturer.

Calibration Steps:

  • Begin with the pH 7.00 buffer solution (neutral).
  • Rinse the electrode with distilled water, and then immerse it in the pH 7.00 solution.
  • Allow the reading to stabilize and adjust the meter to read 7.00 if necessary.Next, calibrate with the pH 4.00 buffer (acidic).
  • Rinse the electrode with distilled water. Rinse and pat dry with a lint-free tissue in between buffers.
  • Rinse the electrode, immerse it in the pH 4.00 solution, and adjust the meter accordingly.

Standard buffer solution

Observation Reading

Trail 1

Trail 2

Trail 3

Buffer 7

7.02

6.98

6.96

Buffer 4

4.03

4.02

4.03

OBSERVATION TABLET : Table: calibration of pH meter

ACCURACY :

observed value should be within +/-5 value of standard pH.

It means that pH 6.95 to 7.05 it would be acceptable for pH 7 and 3.95 to 4.05 it would be acceptable for pH 4.

CONCLUSION :

The pH meter was successfully cleaned and calibrated.

All readings were within acceptable limits, confirming the accuracy of the pH meter.

CLEANING & CALIBRATION OF WEIGHING BALANCE :

1. Introduction :

Provide a brief overview of the Importance of weighing balances in laboratory settings, particularly in QA and QC processes. Discuss how accurate measurements are crucial for experiments, quality control, and regulatory compliance

2. Objective :

Clearly state the aim of the report. For example

  • To perform proper cleaning and maintenance of the weighing balance.
  • To calibrate the weighing Balance to ensure accurate measurements.

3. Materials and Equipment :

List the materials and equipment used during the cleaning and calibration process, Include

  • Weighing balance
  • Cleaning agents (eg, ethanol, distilled water)
  • Soft cloth or lint-free wipes.
  • Calibration weights (traceable to national standards)
  • Calibration certificate (if applicable)

4. Procedure :

  1. Cleaning Procedure :

Preparation :

  • Ensure the balance is turned off and unplugged.
  • Gather all necessary cleaning materials.

Cleaning Steps :

  • Use a soft cloth dampened with distilled water or a mild cleaning agent to wipe the exterior surfaces.
  • Avoid abrasive materials that could scratch the balance.
  • Clean the weighing pan with ethanol and a allow it to dry completely.
  • Check for dust or debris in the balance's internal components and carefully remove it if accessible.
  • Document any issues found during the cleaning.
  1. Calibration Procedure :

Preparation :

  • Turn on the weighing balance and allow it to warm up for at least 30 minutes.
  • Ensure the balance is level using Built-in levelling feet and spirit level.

Calibration Steps :

  • Access the calibration mode on the balance (refer to the manufacturer's manual).
  • Use standard calibration weights that are traceable to national standards.
  • Place the calibration weight on the balance, allowing the reading to stabilize
  • Record the displayed weight and compare it to the actual weight of the calibration
  • weight.
  • Adjust the balance settings if discrepancies exceed acceptable limits.
  • Repeat the process for multiple weights (eg, 50mg, 100mg, and 200mg) to ensure accuracy across the range.
  • Document the calibration results and adjustments made

OBSERVATION TABLE :

  1. Accuracy/Linearity :

Task

Trail 1

Trail 2

Trail 3

50mg

49.45

50.00

51.00

100mg

99.95

102

104

200mg

199.50

200.65

200.75

  1. Repeatability/Uncertanity :

Weight. No

Reading displayed by positive balance

1

20.006

2

20.007

3

20.003

4

20.004

5

20.003

6

20.002

7

20.002

8

20.004

9

20.002

10

20.002

CONCLUSION:

The weighing balance was successfully calibrated using certified standard weights. The balance demonstrates accurate, consistent, and reliable performance and is deemed suitable for use in routine weighing operations..

 

REFERENCES

  1. Gowthami, Gattu Venkateshwarlu, P.Venkateshwarlu, (2014), Calibration and validation of ph meter. Weighing balance  International journal of modern chemistry and applied science.
  2. Gowrishankar D., Abbula K., Bala Souri O., & Sujana K. (2010). “Validation and Calibration of Analytical Instruments.” J Biomed Sci and Res.
  3. https://helpjuice.com/blog/standard-operating-procedure
  4. CH. Aparna and D. Gowrisankar, (2015), A review on calibration of analytical instruments. International journal of pharmaceutical chemical and biological sciences,5(3):572-582.
  5. Yang Wang, Harold E. Van Wart, (1993). Raman and resonance Raman spectroscopy, Methods in Enzymology, Academic Press, Volume 226,ISBN 9780121821272,https
  6. Hobert H. Willard, Lynne L. Merritt Jr., John A. Dean, Frank A. Settle Jr., Instrumental methods of analysis, Seventh edition, CBS publishers and distributors Pvt. Ltd, Page no.- 03 and 05.
  7. Panchumarthy Ravisankar, Ch. Naga Navya1, D. Pravallika , D. Navya Sri, A Review on Step-by-Step Analytical Method Validation, IOSR Journal of pharmacy, October 2015, 7-8.
  8. Francisco Raposo, Evaluation of analytical calibration based on least-squares linear regression for instrumental techniques: A tutorial review, 2015 (167-185).
  9. G. Lavanya, M. Sunil, M.M. Eswarudu, M. Chinna Eswaraiah, K. Harisudha and B. Naga Spandana, Department of Pharmaceutical Analysis, IJPSR, 2013; Vol. 4(4): 1280-1286
  10. Masoom Raza Siddiqui, Zeid A. Alothman, Nafisur Rahman, Analytical techniques in pharmaceutical analysis: Review, 2013, 1409- 1421.

Reference

  1. Gowthami, Gattu Venkateshwarlu, P.Venkateshwarlu, (2014), Calibration and validation of ph meter. Weighing balance  International journal of modern chemistry and applied science.
  2. Gowrishankar D., Abbula K., Bala Souri O., & Sujana K. (2010). “Validation and Calibration of Analytical Instruments.” J Biomed Sci and Res.
  3. https://helpjuice.com/blog/standard-operating-procedure
  4. CH. Aparna and D. Gowrisankar, (2015), A review on calibration of analytical instruments. International journal of pharmaceutical chemical and biological sciences,5(3):572-582.
  5. Yang Wang, Harold E. Van Wart, (1993). Raman and resonance Raman spectroscopy, Methods in Enzymology, Academic Press, Volume 226,ISBN 9780121821272,https
  6. Hobert H. Willard, Lynne L. Merritt Jr., John A. Dean, Frank A. Settle Jr., Instrumental methods of analysis, Seventh edition, CBS publishers and distributors Pvt. Ltd, Page no.- 03 and 05.
  7. Panchumarthy Ravisankar, Ch. Naga Navya1, D. Pravallika , D. Navya Sri, A Review on Step-by-Step Analytical Method Validation, IOSR Journal of pharmacy, October 2015, 7-8.
  8. Francisco Raposo, Evaluation of analytical calibration based on least-squares linear regression for instrumental techniques: A tutorial review, 2015 (167-185).
  9. G. Lavanya, M. Sunil, M.M. Eswarudu, M. Chinna Eswaraiah, K. Harisudha and B. Naga Spandana, Department of Pharmaceutical Analysis, IJPSR, 2013; Vol. 4(4): 1280-1286
  10. Masoom Raza Siddiqui, Zeid A. Alothman, Nafisur Rahman, Analytical techniques in pharmaceutical analysis: Review, 2013, 1409- 1421.

Photo
Shaikh Shahebaz Shaikh Mushtaque
Corresponding author

PRMSS Anuradha College of pharmacy, Chikhli:443201 Maharashtra, India.

Photo
Prashant Waghmode
Co-author

PRMSS Anuradha College of pharmacy, Chikhli:443201 Maharashtra, India.

Photo
Rupesh Gedam
Co-author

PRMSS Anuradha College of pharmacy, Chikhli:443201 Maharashtra, India.

Photo
Rushikesh Payghan
Co-author

PRMSS Anuradha College of pharmacy, Chikhli:443201 Maharashtra, India.

Photo
Rushikesh Suradkar
Co-author

PRMSS Anuradha College of pharmacy, Chikhli:443201 Maharashtra, India.

Photo
Sanket Sardar
Co-author

PRMSS Anuradha College of pharmacy, Chikhli:443201 Maharashtra, India.

Photo
DR. R. H. Kale
Co-author

PRMSS Anuradha College of pharmacy, Chikhli:443201 Maharashtra, India.

Prashant Waghmode, Rupesh Gedam, Rushikesh Payghan, Rushikesh Suradkar, Sanket Sardar, Shaikh Shahebaz Shaikh Mushtaque*, DR. R. H. Kale, A Comprehensive Study on The Calibration Techniques of Ph Meter and Analytical Balances in Laboratory Setting: A Review, Int. J. of Pharm. Sci., 2025, Vol 3, Issue 6, 727-733. https://doi.org/10.5281/zenodo.15595734

More related articles
A Review on Acute Brewery Syndrome...
Tripuramallu Rajithasree, Potti Jessica, Pendyala Keerthana, R. S...
Liquid Liquid Separation And Determination Of Beta...
Edebi N. Vaikosen, Ruth C. Worlu, Samuel J. Bunu, Ebiere Dode, Ma...
Colocasia esculenta (Taro): A Comprehensive Review...
Yogeshwari Mali, Prapti Jadhav, Nikita R. K., Mohazzib Husain, Wa...
Formulation And Evaluation Of Polyherbal Wound Healing Ointment Using Chromolaen...
Sainabi P, Meghna, Thashreefa P M, Sheetal Krishna T, Vyshnavy Devy D K, ...
Evaluation Of Efficacy of Lanthanum Carbonate in Rats with Chronic Kidney Diseas...
Dr. Sushma Ghadigaonkar, S. S. Jadhav, M. G. Ghadigaonkar, ...
Evaluation And Preparation of Herbal Hair Oil ...
Prajakta Vilas Patil, Jagruti Sunil Patil, Kuldip Ravindra Patil, Dr. Gajanan Tulsiram Daphal, Dr. S...
Related Articles
Gallic Acid Nanofiber Loaded in Polymer PVP And Peg by Electrospun...
Dr. Omprakash Bhusnure , Rajashree Biradar, Shivam Vyavahare , Mani Ganesh, Vijayendra Swammy , Hyun...
Recurrent Phenytoin-Induced Gingival and Lip Hyperplasia Following Rat Killer Pa...
G. Georglin Heavenlia , R. S. Atchayavarshini, B. Deepasree, Fathima Juhaina. M. Abdul Khader, Dr. V...
Formulation And Evaluation Of Herbal Cream For Management Of Varicose Veins...
Kartiki Dnyaneshwar More , Pankaj Vishnu Vyawhare , ...
A Review: Gastroretentive Drug Delivery Systems...
Dr. Jeevan R. Rajguru, Gauri R. Dhamale, Harshali L. Shelke, Priti B. Shinde, Rohini R. Khedkar, Pre...
A Review on Acute Brewery Syndrome...
Tripuramallu Rajithasree, Potti Jessica, Pendyala Keerthana, R. Sakshi, Golla Akhila, Chandrasekhara...
More related articles
A Review on Acute Brewery Syndrome...
Tripuramallu Rajithasree, Potti Jessica, Pendyala Keerthana, R. Sakshi, Golla Akhila, Chandrasekhara...
Liquid Liquid Separation And Determination Of Betamethasone, Clotrimazole And Ne...
Edebi N. Vaikosen, Ruth C. Worlu, Samuel J. Bunu, Ebiere Dode, Mary Doctor, ...
Colocasia esculenta (Taro): A Comprehensive Review of Phytochemistry, Pharmacolo...
Yogeshwari Mali, Prapti Jadhav, Nikita R. K., Mohazzib Husain, Wadulkar R. D., Satpute K. L., ...
A Review on Acute Brewery Syndrome...
Tripuramallu Rajithasree, Potti Jessica, Pendyala Keerthana, R. Sakshi, Golla Akhila, Chandrasekhara...
Liquid Liquid Separation And Determination Of Betamethasone, Clotrimazole And Ne...
Edebi N. Vaikosen, Ruth C. Worlu, Samuel J. Bunu, Ebiere Dode, Mary Doctor, ...
Colocasia esculenta (Taro): A Comprehensive Review of Phytochemistry, Pharmacolo...
Yogeshwari Mali, Prapti Jadhav, Nikita R. K., Mohazzib Husain, Wadulkar R. D., Satpute K. L., ...